2024/3/13
Dustproofing test for all kinds of prototypes
Adaptation standards
IEC 60529, IP5X/6X
GB/T 4208, IP5X/6X
ISO 20653, IP5X/6X
GB/T 30038, IP5X/6X
IEC 60068-2-68, La2
GB/T 2423.37, La2
JIS D 0207
IEC 60598-1, IP5X/6X
GB 7000.1, IP5X/6X
GB/T 10485, 防尘试验
IEC 60034-5, IP5X/6X
GB/T 4942.1, IP5X/6X
SAE J 575, Dust Exposure Test
Specificities
- The inner liner of the test chamber is made of high-quality, corrosion-resistant stainlesssteel with frequency conversion to control the air flow rate inside the chamber. |
|
- The blower is used to realize the circulation of talcum powder in the box.
- The blower is controlled by a frequency converter, which can be set at different frequencies to achieve different air speeds.
- The dust entering the blower is pre-screened.
- A vibrator is used to vibrate the dust scattered on the walls of the chamber down to the screen.
- According to the test requirements into the appropriate amount of dust, not more than 4kg
- The equipment is equipped with a sample pumping system, when you need to use the sample in the box and the pumping interface connected to the negative pressure.
- The pumping system consists of a vacuum pump, a dust filter, a differential pressure sensor, an air flow meter and a flow control valve.
- After calculating the pumping flow rate according to the test requirements, the required flow rate can be achieved by adjusting the flow regulating valve.
- By blowing air or fan (blowing air is used in this unit), the dust is lifted up at regular intervals and distributed evenly in the box.
- Different levels of dust concentration can be realized by setting different blowing pressures.
- The test temperature is 20℃±15℃ and the test humidity is 45% RH~85% RH.
- There is a heating plate attached to the wall of the box, and the temperature inside the box can be set to achieve the appropriate temperature and humidity.
- Temperature and humidity sensors are installed inside the chamber to collect the temperature and relative humidity.
系统参数与操作
|
① The main signals and alarm triggers of the equipment are displayed, such as gas source pressure, phase sequence of power supply and door signal. ② test chamber set temperature, real-time temperature and real-time humidity curve display ③ Digital display of test chamber set temperature, real-time temperature, real-time humidity and wind speed. ④ Display of operation mode, relative pressure and flow rate of vacuum circuit ⑤ Timing display of vibration cycle and dust blowing cycle. ⑥ Display of test progress ⑦ Equipment operation information prompts
⑧ Protective operation and stop touch keys, each operation needs to be confirmed.
|
|
①Test Chamber Temperature Setting ② Operation mode selection: C1/C2/C3 ③ Wind speed setting: max. 25m/s ④ Test total time setting, maximum value 99999 hours ⑤ Vibration cycle and time setting ⑥ Dust blowing cycle time setting ⑦ If equipped with serial port communication, you can set the communication baud rate and node number.
⑧ Each parameter modification needs to be saved before taking effect.
|
Shanghai Acutech Automation Testing Technology Co., Ltd. Copyright 沪ICP备18019040号-1 {Disclaimer}